이경중,송향봉 대한의용생체공학회 1987 의공학회지 Vol.8 No.2
This paper describes the design of a real time arrhythmia detection system using five variables : heart rate, QS width, morphology, alls deviation, and ST segment. To detect individual variables for four patients, we designed multi-processor system. The results of the analysis derived from simulators and 50 patients are compared with the medical diagnoses. The results show that the analysis was able to detect irregularly occuring arrhythmia which does not show up in routine medical examination.