Blood pump analysis process includes both mechanical and bio-mechanical aspects. Since a blood pump is a mechanical device, it has to be mechanically efficient. On the other hand, blood pumps function is sensitively related to the blood recirculation;...
Blood pump analysis process includes both mechanical and bio-mechanical aspects. Since a blood pump is a mechanical device, it has to be mechanically efficient. On the other hand, blood pumps function is sensitively related to the blood recirculation; hence, bio-factors such as hemolysis and thrombosis become important. This paper numerically investigates the mechanical and bio-mechanical performances of the Rotaflow in the extracorporeal membrane oxygenation(ECMO), Ventricular Assist Device(VAD), and full-load conditions. The operational conditions are defined as(400[mmHg], 5[L/min.]), (100[mmHg], 3[L/min.]), and (600[mmHg], 10[L/min.]) for ECMO, VAD, and full-load conditions, respectively. The results are presented and analyzed from the mechanical aspect via performance curves, and from bio-mechanical aspect via focusing on hemolytic characteristics. Regions of top and bottom cavities show recirculation in both ECMO and VAD condtions. In addition, Eulerian-based calculation of modified index of hemolysis(MIH) has been investigated. The results demonstrate that the VAD condition has the least risk of hemolysis among the others, while the full-load condition has the highest risk.