      Evaluation of Automated Platelet Aggregation Test Using a Sysmex CS-5100 Analyzer

      한글로보기

      https://www.riss.kr/link?id=A106829934

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract)

      Background: The platelet aggregation test is widely used to measure antiplatelet therapy response and to detect platelet function disorders. CS-5100 (Sysmex Co., Japan) is a recently introduced coagulation analyzer that can also measure platelet aggre...

      Background: The platelet aggregation test is widely used to measure antiplatelet therapy response and to detect platelet function disorders. CS-5100 (Sysmex Co., Japan) is a recently introduced coagulation analyzer that can also measure platelet aggregation. We evaluated the performance of CS-5100 in the platelet aggregation test for use in clinical laboratories. Methods: We investigated the precision, stability, dilution test, and correlation of CS-5100 with a traditional aggregometer. Precision was tested using normal and patient samples. Stability was assessed over 5 different time points for 8 hours. The dilution test was performed with normal samples using ADP agonists. We tested correlations between the results of Chrono-log aggregometer (Chrono-Log Co., U.S.A) and CS-5100 using 10 samples with 5 agonists each. We also calculated the reference range of 5 agonists using 22-30 normal samples. Results: The coefficients of variation using normal samples were 7.45% and 3.27% for ADP and arachidonic acid, respectively. Stability was maintained for up to 2 hours in most samples. Dilution tests showed similar results until reaching a dilution factor of 2. Strong correlations of the results between Chrono-log and CS-5100 were found, except for ristocetin. The reference ranges of 5 reagents in CS-5100 were similar to those obtained with the Chrono-log aggregometer. Conclusions: The performance of CS-5100 in platelet aggregation tests showed reliable results compared to a traditional aggregometer. CS5100 can perform coagulation test and platelet aggregation test, simultaneously. Thus, CS-5100 can enable cost saving and reduce turn-aroundtime by reducing the inspection time.

      더보기

      참고문헌 (Reference)

      1 Lawrie AS, "The automation of routine light transmission platelet aggregation" 36 : 431-438, 2014

      2 Born GV, "The aggregation of blood platelets" 168 : 178-195, 1963

      3 Mukaka MM, "Statistics corner : A guide to appropriate use of correlation coefficient in medical research" 24 : 69-71, 2012

      4 Cattaneo M, "Results of a worldwide survey on the assessment of platelet function by light transmission aggregometry : a report from the platelet physiology subcommittee of the SSC of the ISTH" 7 : 1029-, 2009

      5 Cattaneo M, "Recommendations for the standardization of light transmission aggregometry : a consensus of the working party from the platelet physiology subcommittee of SSC/ISTH" 11 : 1183-1189, 2013

      6 Harrison P, "Platelet function analysis" 19 : 111-123, 2005

      7 Clinical and Laboratory Standard Institute, "Measurement procedure comparison and bias estimation using patient samples; Approved guideline" Clinical and Laboratory Standard Institute 2013

      8 Sakayori T, "Introduction and evaluation of light transmission platelet aggregation method on the Sysmex CS-series automated coagulation analyzer" 26 : 1-10, 2016

      9 Taylor R, "Interpretation of the correlation coefficient : a basic review" 6 : 35-39, 1990

      10 Miller CH, "Gender, race and diet affect platelet function tests in normal subjects, contributing to a high rate of abnormal results" 165 : 842-853, 2014

      1 Lawrie AS, "The automation of routine light transmission platelet aggregation" 36 : 431-438, 2014

      2 Born GV, "The aggregation of blood platelets" 168 : 178-195, 1963

      3 Mukaka MM, "Statistics corner : A guide to appropriate use of correlation coefficient in medical research" 24 : 69-71, 2012

      4 Cattaneo M, "Results of a worldwide survey on the assessment of platelet function by light transmission aggregometry : a report from the platelet physiology subcommittee of the SSC of the ISTH" 7 : 1029-, 2009

      5 Cattaneo M, "Recommendations for the standardization of light transmission aggregometry : a consensus of the working party from the platelet physiology subcommittee of SSC/ISTH" 11 : 1183-1189, 2013

      6 Harrison P, "Platelet function analysis" 19 : 111-123, 2005

      7 Clinical and Laboratory Standard Institute, "Measurement procedure comparison and bias estimation using patient samples; Approved guideline" Clinical and Laboratory Standard Institute 2013

      8 Sakayori T, "Introduction and evaluation of light transmission platelet aggregation method on the Sysmex CS-series automated coagulation analyzer" 26 : 1-10, 2016

      9 Taylor R, "Interpretation of the correlation coefficient : a basic review" 6 : 35-39, 1990

      10 Miller CH, "Gender, race and diet affect platelet function tests in normal subjects, contributing to a high rate of abnormal results" 165 : 842-853, 2014

      11 Clinical and Laboratory Standard Institute, "Evaluation of precision of quantitative measurement procedures; Approved guideline" Clinical and Laboratory Standard Institute 2014

      12 Ling LQ, "Evaluation of an automated light transmission aggregometry" 28 : 712-719, 2017

      13 Clinical and Laboratory Standard Institute, "Defining, establishing, and verifying reference intervals in the clinical laboratory; Approved guideline" Clinical and Laboratory Standard Institute 2010

      14 Gale AJ, "Continuing education course #2 : Current understanding of hemostasis" 39 : 273-280, 2011

      15 Nurden AT, "Congenital platelet disorders and understanding of platelet function" 165 : 165-178, 2014

      16 Hayward CP, "An evaluation of methods for determining reference intervals for light transmission platelet aggregation tests on samples with normal or reduced platelet counts" 100 : 134-145, 2008

      17 Konkle BA, "Acquired disorders of platelet function" 2011 : 391-396, 2011

      더보기

      분석정보

      View

      상세정보조회

      0

      Usage

      원문다운로드

      0

      대출신청

      0

      복사신청

      0

      EDDS신청

      0

      동일 주제 내 활용도 TOP

      더보기

      주제

      연도별 연구동향

      연도별 활용동향

      연관논문

      연구자 네트워크맵

      공동연구자 (7)

      유사연구자 (20) 활용도상위20명

      인용정보 인용지수 설명보기

      학술지 이력

      학술지 이력
      연월일 이력구분 이력상세 등재구분
      2027 평가예정 재인증평가 신청대상 (재인증)
      2021-01-01 평가 등재학술지 유지 (재인증) KCI등재
      2018-01-01 평가 등재학술지 유지 (등재유지) KCI등재
      2015-01-01 평가 등재학술지 선정 (계속평가) KCI등재
      2013-01-01 평가 등재후보학술지 유지 (기타) KCI등재후보
      2012-01-01 평가 등재후보학술지 유지 (기타) KCI등재후보
      2011-01-01 평가 등재후보 1차 PASS (등재후보1차) KCI등재후보
      2011-01-01 평가 학술지 분리 (기타) KCI등재후보
      더보기

      학술지 인용정보

      학술지 인용정보
      기준연도 WOS-KCI 통합IF(2년) KCIF(2년) KCIF(3년)
      2016 0.16 0.16 0.11
      KCIF(4년) KCIF(5년) 중심성지수(3년) 즉시성지수
      0.09 0.09 0.306 0.05
      더보기

      이 자료와 함께 이용한 RISS 자료

      나만을 위한 추천자료

      해외이동버튼