<P>We report a rare case of vascular injury secondary to a damaged Hot Shears™ tip cover. Two 1 mm holes in the tip cover resulted in perforations in the obturator and external iliac veins during pelvic node dissection. Bleeding was controlled...
http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
https://www.riss.kr/link?id=A107572893
2011
-
SCI,SCIE,SCOPUS,KCI등재
학술저널
365-368(4쪽)
0
상세조회0
다운로드다국어 초록 (Multilingual Abstract)
<P>We report a rare case of vascular injury secondary to a damaged Hot Shears™ tip cover. Two 1 mm holes in the tip cover resulted in perforations in the obturator and external iliac veins during pelvic node dissection. Bleeding was controlled...
<P>We report a rare case of vascular injury secondary to a damaged Hot Shears™ tip cover. Two 1 mm holes in the tip cover resulted in perforations in the obturator and external iliac veins during pelvic node dissection. Bleeding was controlled with bipolar coagulation and a 5 mm metal clip in the obturator and iliac vein, respectively. The rest of the procedure was completed uneventfully. Frequent integrity assessment of this accessory is necessary. Its function is important in order to carry out safe dissection in proximity to delicate structures. When injuries arise from areas not directly involved in the dissection, immediate inspection of the instruments should be mandatory.</P>