      콘빔전산화단층촬영에서 노출 조건에 따른 화질 유지 및 선량 감소에 대한 평가 = Evaluation of Dose Reduction and Maintaining Image Quality according to Exposure Factors of ConeBeam Computed Tomography

      한글로보기

      https://www.riss.kr/link?id=A107021773

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract)

      This research aims at suggesting exposure condition that shows maintaining the value of the physical image quality factor by decreasing tube voltage and tube current from the standard exposure condition(80 kV, 7 mA) of a CBCT apparatus. To measure th...

      This research aims at suggesting exposure condition that shows maintaining the value of the physical image quality factor by decreasing tube voltage and tube current from the standard exposure condition(80 kV, 7 mA) of a CBCT apparatus. To measure the value of the physical image quality factor, modular transfer function(MTF) was analyzed and dose-area product(DAP) was used for the measurement of exposure dose. CBCT images of a Sedentex IQ phantom were obtained under 15 exposure conditions of different combination of tube voltage(80, 78, 76 kV) and tube current(7, 6, 5, 4, 3 mA) and MTF 10 was calculated under each exposure conditions. There were no significant differences in MTF 10 under 80 kV-6 mA, 80 kV-5 mA exposure conditions in comparison with standard exposure condition. Based on the results of this research, 80 kV-5 mA condition are expected to be able to reduce exposure dose with maintaining the value of the physical image quality factor of the standard exposure condition.

      더보기

      국문 초록 (Abstract)

      본 연구는 CBCT 장비의 표준 노출 조건(80 kV, 7 mA)을 기준으로 관전압과 관전류를 낮추어 촬영하면서 표준 노출 조건의 물리적 화질 요소값을 유지할 수 있는 노출 조건의 제시를 목표로 두고...

      본 연구는 CBCT 장비의 표준 노출 조건(80 kV, 7 mA)을 기준으로 관전압과 관전류를 낮추어 촬영하면서 표준 노출 조건의 물리적 화질 요소값을 유지할 수 있는 노출 조건의 제시를 목표로 두고자한다. 영상의 물리적 화질 요소값의 측정을 위해 변조전달함수(MTF)가 분석되었고 선량 측정을 위해 선량-면적 곱(DAP)을 이용하였다. 관전압(80, 78, 76 kV)과 관전류(7, 6, 5, 4, 3 mA)의 15가지 조합의 노출 조건에서 Sedentex IQ 팬텀 (Leeds Test Objects Ltd., Boroughbridge, UK)의 CBCT 영상을 얻었고 MTF 10이 각 조건에서 계산 되었다. 표준 노출 조건과 비교시 80 kV-6 mA, 80 kV-5 mA 노출 조건은 MTF 10에 있어 유의한 차이를 보이지 않았기에 본 연구에서 사용된 CBCT 장비의 경우, 80 kV-5 mA로 낮춘 노출 조건에서 물리적 화질 요소값을 유지하면서 선량을 감소시킬 수 있는 것으로 판단되어진다.

      더보기

      참고문헌 (Reference)

      1 "http://www.sedentexct.eu/files/radiation_protection_172. pdf"

      2 Jin-Woo Choi, "Relationship between physical factors and subjective image quality of cone-beam computed tomography images according to diagnostic task" Elsevier BV 119 (119): 357-365, 2015

      3 박하나, "Optimization of exposure parameters and relationship between subjective and technical image quality in cone-beam computed tomography" 대한영상치의학회 49 (49): 139-151, 2019

      4 Satomi Nakahara, "One-year analysis of Elekta CBCT image quality using NPS and MTF" Wiley 17 (17): 211-222, 2016

      5 Yoshihiro Ozaki, "Location dependency of the spatial resolution of cone beam computed tomography for dental use" Elsevier BV 116 (116): 648-655, 2013

      6 Sara M Elkhateeb, "Image quality assessment of clinically-applied CBCT protocols using a QAT phantom" British Institute of Radiology 45 (45): 20160075-, 2016

      7 최진우, "Factors affecting modulation transfer function measurements in cone-beam computed tomographic images" 대한영상치의학회 49 (49): 131-137, 2019

      8 Sara Lofthag-Hansen, "Evaluation of subjective image quality in relation to diagnostic task for cone beam computed tomography with different fields of view" Elsevier BV 80 (80): 483-488, 2011

      9 A. Suomalainen, "Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners" 38 (38): 367-378, 2009

      10 Korea Centers for Disease Control and Prevention, "Diagnostic Reference Level guideline – Dental Radiography"

      1 "http://www.sedentexct.eu/files/radiation_protection_172. pdf"

      2 Jin-Woo Choi, "Relationship between physical factors and subjective image quality of cone-beam computed tomography images according to diagnostic task" Elsevier BV 119 (119): 357-365, 2015

      3 박하나, "Optimization of exposure parameters and relationship between subjective and technical image quality in cone-beam computed tomography" 대한영상치의학회 49 (49): 139-151, 2019

      4 Satomi Nakahara, "One-year analysis of Elekta CBCT image quality using NPS and MTF" Wiley 17 (17): 211-222, 2016

      5 Yoshihiro Ozaki, "Location dependency of the spatial resolution of cone beam computed tomography for dental use" Elsevier BV 116 (116): 648-655, 2013

      6 Sara M Elkhateeb, "Image quality assessment of clinically-applied CBCT protocols using a QAT phantom" British Institute of Radiology 45 (45): 20160075-, 2016

      7 최진우, "Factors affecting modulation transfer function measurements in cone-beam computed tomographic images" 대한영상치의학회 49 (49): 131-137, 2019

      8 Sara Lofthag-Hansen, "Evaluation of subjective image quality in relation to diagnostic task for cone beam computed tomography with different fields of view" Elsevier BV 80 (80): 483-488, 2011

      9 A. Suomalainen, "Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners" 38 (38): 367-378, 2009

      10 Korea Centers for Disease Control and Prevention, "Diagnostic Reference Level guideline – Dental Radiography"

      11 Y. Arai, "Development of a compact computed tomographic apparatus for dental use" 28 (28): 245-248, 1999

      12 김동수, "Conversion coefficients for the estimation of effective dose in cone-beam CT" 대한영상치의학회 44 (44): 21-29, 2014

      13 Ruben Pauwels, "Comparison of spatial and contrast resolution for cone-beam computed tomography scanners" Elsevier BV 114 (114): 127-135, 2012

      14 C. M. Ziegler, "Clinical indications for digital volume tomography in oral and maxillofacial surgery" 31 (31): 126-130, 2002

      15 안창현, "CBCT의 선택과 처방" 대한치과의사협회 52 (52): 132-138, 2014

      16 최용석, "CBCT 검사법의 정도관리 및 선량" 대한치과의사협회 52 (52): 153-163, 2014

      17 Munetaka Naitoh, "Assessment of Three-dimensional X-ray Images: Reconstruction From Conventional Tomograms, Compact Computerized Tomography Images, and Multislice Helical Computerized Tomography Images" American Academy of Implant Dentistry 31 (31): 234-241, 2005

      더보기

      분석정보

      View

      상세정보조회

      0

      Usage

      원문다운로드

      0

      대출신청

      0

      복사신청

      0

      EDDS신청

      0

      동일 주제 내 활용도 TOP

      더보기

      주제

      연도별 연구동향

      연도별 활용동향

      연관논문

      연구자 네트워크맵

      공동연구자 (7)

      유사연구자 (20) 활용도상위20명

      인용정보 인용지수 설명보기

      학술지 이력

      학술지 이력
      연월일 이력구분 이력상세 등재구분
      2026 평가예정 재인증평가 신청대상 (재인증)
      2020-01-01 평가 등재학술지 유지 (재인증) KCI등재
      2017-01-01 평가 등재학술지 선정 (계속평가) KCI등재
      2016-01-01 평가 등재후보학술지 유지 (계속평가) KCI등재후보
      2015-01-01 평가 등재후보학술지 유지 (계속평가) KCI등재후보
      2013-01-01 평가 등재후보학술지 유지 (기타) KCI등재후보
      2012-01-01 평가 등재후보학술지 유지 (기타) KCI등재후보
      2011-02-28 학술지명변경 한글명 : 한국방사선학회 논문지 -> 한국방사선학회논문지 KCI등재후보
      2010-01-01 평가 등재후보학술지 선정 (신규평가) KCI등재후보
      2008-01-24 학회명변경 한글명 : 방사선학회 -> 한국방사선학회
      영문명 : The Society of Radiology -> The Korea Society of Radiology
      더보기

      학술지 인용정보

      학술지 인용정보
      기준연도 WOS-KCI 통합IF(2년) KCIF(2년) KCIF(3년)
      2016 0.28 0.28 0.36
      KCIF(4년) KCIF(5년) 중심성지수(3년) 즉시성지수
      0.37 0.37 0.452 0.05
      더보기

      이 자료와 함께 이용한 RISS 자료

      나만을 위한 추천자료

      해외이동버튼