<P>A new medium vacuum primary standard using the static expansion method was developed in Korea Research Institute of Standards and Technology. We compared the system with an ultrasonic interferometer manometer using two capacitance diaphragm g...
http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
https://www.riss.kr/link?id=A107691665
2005
-
SCOPUS,SCIE
학술저널
173-175(3쪽)
0
상세조회0
다운로드다국어 초록 (Multilingual Abstract)
<P>A new medium vacuum primary standard using the static expansion method was developed in Korea Research Institute of Standards and Technology. We compared the system with an ultrasonic interferometer manometer using two capacitance diaphragm g...
<P>A new medium vacuum primary standard using the static expansion method was developed in Korea Research Institute of Standards and Technology. We compared the system with an ultrasonic interferometer manometer using two capacitance diaphragm gauges with full-scale ranges of 133 Pa and 1333 Pa. The results showed that the two standards are coincident with each other within the range of uncertainty at calibrated pressures 3 Pa to 100 Pa.</P>