      5MHz 초과 초음파자극기에 대한 IEC 61689 규격 적용의 유효성에 관한 연구 = A Study on Effectiveness of Application of the IEC 61689 Standard to Ultrasonic Physiotherapy Systems with Frequency Range over 5MHz

      한글로보기

      https://www.riss.kr/link?id=A101120874

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract)

      Ultrasonic physiotherapy systems should comply with IEC 60601-2-5(particular requirements for safety of ultrasonic physiotherapy equipment) standard for safety of patients and clinical performances. One of the most important parameters in the standard...

      Ultrasonic physiotherapy systems should comply with IEC 60601-2-5(particular requirements for safety of ultrasonic physiotherapy equipment) standard for safety of patients and clinical performances. One of the most important parameters in the standard is the effective radiating area ($A_{ER}$). It has to be measured in accordance with IEC 61689 (field specifications and methods of measurement in the frequency range 0.5MHz to 5MHz). Typical ultrasonic physiotherapy system works in the frequency range 1MHz ~ 3MHz to comply with the IEC 61689. However, ultrasonic physiotherapy system using frequencies over 5MHz is out of the IEC 61689's scope. That is, even if such ultrasonic physiotherapy systems are developed by demands of the market, there is no standard to apply. It is the motivation for this study. Whereas there are other parameters to be considered, this study focuses on the effective radiating area and shows effectiveness of applying IEC 61689 in measuring effective radiating area of ultrasonic physiotherapy systems using frequency range over 5MHz by comparing the results of computer simulation and experiment. Results of this study shows that applying the IEC 61689 standard to ultrasonic physiotherapy system using frequency range over 5MHz is possible.

      더보기

      참고문헌 (Reference)

      1 "Regulations for classification of medical devices, No. 2014-110"

      2 "Particular requirements for the safety of ultrasonic physiotherapy equipment"

      3 Andrè V. Alvarenga, "Metrological aspects on therapeutic ultrasound parameters: effective radiating area and non-uniformity ration" 3 : 643-649, 2010

      4 Mario-Ibrahin Gutièrrez, "Methods for Characterization of Physiotherapy Ultrasonic Transducers"

      5 "IEC 61161 Radiation force balances and performance requirements"

      6 Rinaldo Guirro, "Evaluation of the acoustic intensity of new ultrasound therapy equipment" 39 : 553-557, 2002

      7 Raejoon Park, "Electrical Treatment" Hyunmoonsa 434-, 2009

      1 "Regulations for classification of medical devices, No. 2014-110"

      2 "Particular requirements for the safety of ultrasonic physiotherapy equipment"

      3 Andrè V. Alvarenga, "Metrological aspects on therapeutic ultrasound parameters: effective radiating area and non-uniformity ration" 3 : 643-649, 2010

      4 Mario-Ibrahin Gutièrrez, "Methods for Characterization of Physiotherapy Ultrasonic Transducers"

      5 "IEC 61161 Radiation force balances and performance requirements"

      6 Rinaldo Guirro, "Evaluation of the acoustic intensity of new ultrasound therapy equipment" 39 : 553-557, 2002

      7 Raejoon Park, "Electrical Treatment" Hyunmoonsa 434-, 2009

      더보기

      분석정보

      View

      상세정보조회

      0

      Usage

      원문다운로드

      0

      대출신청

      0

      복사신청

      0

      EDDS신청

      0

      동일 주제 내 활용도 TOP

      더보기

      주제

      연도별 연구동향

      연도별 활용동향

      연관논문

      연구자 네트워크맵

      공동연구자 (7)

      유사연구자 (20) 활용도상위20명

      인용정보 인용지수 설명보기

      학술지 이력

      학술지 이력
      연월일 이력구분 이력상세 등재구분
      2027 평가예정 재인증평가 신청대상 (재인증)
      2021-01-01 평가 등재학술지 유지 (재인증) KCI등재
      2018-01-01 평가 등재학술지 선정 (계속평가) KCI등재
      2017-12-01 평가 등재후보로 하락 (계속평가) KCI등재후보
      2013-01-01 평가 등재 1차 FAIL (등재유지) KCI등재
      2010-01-01 평가 등재학술지 유지 (등재유지) KCI등재
      2008-01-01 평가 등재학술지 유지 (등재유지) KCI등재
      2005-10-06 학술지명변경 외국어명 : 미등록 -> Joural of Biomedical Engineering Research KCI등재
      2005-01-01 평가 등재학술지 선정 (등재후보2차) KCI등재
      2004-01-01 평가 등재후보 1차 PASS (등재후보1차) KCI등재후보
      2003-01-01 평가 등재후보학술지 유지 (등재후보1차) KCI등재후보
      2002-01-01 평가 등재후보학술지 유지 (등재후보1차) KCI등재후보
      1999-07-01 평가 등재후보학술지 선정 (신규평가) KCI등재후보
      더보기

      학술지 인용정보

      학술지 인용정보
      기준연도 WOS-KCI 통합IF(2년) KCIF(2년) KCIF(3년)
      2016 0.08 0.08 0.12
      KCIF(4년) KCIF(5년) 중심성지수(3년) 즉시성지수
      0.11 0.09 0.307 0.04
      더보기

      이 자료와 함께 이용한 RISS 자료

      나만을 위한 추천자료

      해외이동버튼