      뇌정위적 방사선수술 시 콜리메이터 크기 변화에 따른 검출기 의존성 평가 = Evaluation of Detector Dependency on Collimator in SRS: Compared Detectors; CC01, CC13, SFD

      한글로보기

      https://www.riss.kr/link?id=A101423527

      • 0

        상세조회
      • 0

        다운로드
      서지정보 열기
      • 내보내기
      • 내책장담기
      • 공유하기
      • 오류접수

      부가정보

      다국어 초록 (Multilingual Abstract)

      Purpose: To evaluate the detector dependency in the various collimator size for Stereotactic Radiosugery (SRS). Materials and Methods: This study was performed with 6 MV photon beam (Varian 21EX, Varian, US) and the measurement detectors are used by i...

      Purpose: To evaluate the detector dependency in the various collimator size for Stereotactic Radiosugery (SRS). Materials and Methods: This study was performed with 6 MV photon beam (Varian 21EX, Varian, US) and the measurement detectors are used by ion chamber CC01, CC13 (Wellhofer, Germany) and stereotactic diode detector (SFD, Wellhofer, Germany). SRS collimator size was used by ${\varphi}$5, 10, 20, 30 mm (Brain Lab, Germany). Percentage depth dose (PDD) was measured at SSD 100 cm and field size 10×10 cm from individual detectors. Ouput factor was measured by using same setup of PDD and with maximum dose depth. Data was normalized at field size $10{\times}10\;cm$. Beam profile was measured at SSD 100 cm in SRS collimator ${\varphi}$10, 30 mm and field $10{\times}10\;cm$ and a comparison of FWHM (full width half maximum), penumbra width (20~80%). Results: The CC13 detector was overestimated 16% than other detectors from the PDD in the 5 mm collimator. Output factors were underestimated CC01 28%, CC13 72% in the 5 mm collimator and CC01 9.6%, CC13 25% in the 10 mm collimator than the SFD. Maximum difference was 3% at the FWHM of the dose profile in the 10 mm collimator and difference of the 30 mm collimator was 0% at the FWHM. Penumbra width was increased CC01 122%, CC13 194% in the 10 mm collimator and CC01 68%, CC13 185% in the 30 mm collimator than the SFD. Conclusion: It is very important for accurate dosimetry to select a detector in small field. The SFD was considered with the most accurate dosimeter for small collimator dosimetry in this study.

      더보기

      분석정보

      View

      상세정보조회

      0

      Usage

      원문다운로드

      0

      대출신청

      0

      복사신청

      0

      EDDS신청

      0

      동일 주제 내 활용도 TOP

      더보기

      주제

      연도별 연구동향

      연도별 활용동향

      연관논문

      연구자 네트워크맵

      공동연구자 (7)

      유사연구자 (20) 활용도상위20명

      이 자료와 함께 이용한 RISS 자료

      나만을 위한 추천자료

      해외이동버튼