It was the purpose of this study to analyze and compare the mechanical properties of orthodontic open coil springs. Four variable facters were presented-wire diameter (.008〃, .009〃, .01〃), lumen size (.030〃, 032〃, .036〃), arch wire size an...
http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
https://www.riss.kr/link?id=A19697650
1992
Korean
515.05
학술저널
265-278(14쪽)
0
상세조회0
다운로드다국어 초록 (Multilingual Abstract)
It was the purpose of this study to analyze and compare the mechanical properties of orthodontic open coil springs. Four variable facters were presented-wire diameter (.008〃, .009〃, .01〃), lumen size (.030〃, 032〃, .036〃), arch wire size an...
It was the purpose of this study to analyze and compare the mechanical properties of orthodontic open coil springs. Four variable facters were presented-wire diameter (.008〃, .009〃, .01〃), lumen size (.030〃, 032〃, .036〃), arch wire size and shape (.016〃found, .016〃×.022〃rect.) and alloy type (HiT Ⅱ, Elgiloy, Sentalloy).
The total 104 specimens were divided into 13 groups, and compression test was performed on an Instron test machine. The load-deflection curve of each open coil spring was obtained, from which, the load-deflection relations, stifnesses, percent recoveries were computed statistically.
The results were obtained as follows
1. When the lumen size of the coil spring remained constant, stiffness and percent recovery increased as the wire diameter increased.
2. When the wire diameter of the coil spring remained constant, stiffness and percent recovery decreased as the lumen size increased.
3. The effect of size and shape of arch wire on the coil spring was not statistically significant.
4. In alloy types, stiffness was the greatest in HiT Ⅱ(55.21), Elgiloy (42.61) and Sentalloy (7.74) in that order. Sentalloy exhibited superior percent recovery and long range of action.