    Reproducibility and diagnostic performance of the vascular index of superb microvascular imaging in real-time breast ultrasonography for evaluating breast masses

    한글로보기

    https://www.riss.kr/link?id=A107591399

    • 0

      상세조회
    • 0

      다운로드
    서지정보 열기
    • 내보내기
    • 내책장담기
    • 공유하기
      • URL 복사
    • 오류접수
    인용문이 복사되었습니다.

    부가정보

    다국어 초록 (Multilingual Abstract) kakao i 다국어 번역

    Purpose: This study aimed to evaluate the reproducibility and diagnostic performance of a quantitative parameter of superb microvascular imaging (SMI) in real-time breast ultrasonography (US) for differentiating benign from malignant breast masses.
    Methods: Eighty-seven breast masses in 75 patients who underwent both B-mode US and SMI before US-guided core needle biopsy were included in this study. Two radiologists performed B-mode US and measured the vascular index (VI) of SMI respectively for each lesion in real time.
    Intraobserver and interobserver agreements were analyzed for the VI of SMI. The diagnostic performance of B-mode US using the Breast Imaging Reporting and Database System lexicon and combined use with the VI of SMI was evaluated compared to pathology.
    Results: The median VI of malignant masses (n=32) was significantly higher than that of benign masses (n=55) (7.6% and 2.6%, respectively; P<0.001). The intraobserver agreement for VI was excellent regardless of the pathology, size, or depth of the lesion. The interobserver agreement for VI was excellent regardless of the presence of a measurement interval. The interobserver agreement for the final diagnostic decision was improved by combining B-mode US and VI (κ=0.883) in comparison with B-mode US only (κ=0.617). Adding VI led to significant improvements in the specificity (87.2% vs. 52.7%, 83.6% vs. 49.0%), accuracy (89.7% vs.
    69.3%, 84.0% vs. 65.9%) and positive predictive value (81.5% vs. 55.1%, 75.6% vs. 52.6%) of B-mode US for both observers compared with B-mode US alone (all, P=0.001).
    Conclusion: The VI of SMI for real-time breast US is highly reproducible and leads to improved diagnostic performance for differentiating between benign and malignant breast lesions in combination with B-mode US.
    번역하기

    Purpose: This study aimed to evaluate the reproducibility and diagnostic performance of a quantitative parameter of superb microvascular imaging (SMI) in real-time breast ultrasonography (US) for differentiating benign from malignant breast masses. Me...

    Purpose: This study aimed to evaluate the reproducibility and diagnostic performance of a quantitative parameter of superb microvascular imaging (SMI) in real-time breast ultrasonography (US) for differentiating benign from malignant breast masses.
    Methods: Eighty-seven breast masses in 75 patients who underwent both B-mode US and SMI before US-guided core needle biopsy were included in this study. Two radiologists performed B-mode US and measured the vascular index (VI) of SMI respectively for each lesion in real time.
    Intraobserver and interobserver agreements were analyzed for the VI of SMI. The diagnostic performance of B-mode US using the Breast Imaging Reporting and Database System lexicon and combined use with the VI of SMI was evaluated compared to pathology.
    Results: The median VI of malignant masses (n=32) was significantly higher than that of benign masses (n=55) (7.6% and 2.6%, respectively; P<0.001). The intraobserver agreement for VI was excellent regardless of the pathology, size, or depth of the lesion. The interobserver agreement for VI was excellent regardless of the presence of a measurement interval. The interobserver agreement for the final diagnostic decision was improved by combining B-mode US and VI (κ=0.883) in comparison with B-mode US only (κ=0.617). Adding VI led to significant improvements in the specificity (87.2% vs. 52.7%, 83.6% vs. 49.0%), accuracy (89.7% vs.
    69.3%, 84.0% vs. 65.9%) and positive predictive value (81.5% vs. 55.1%, 75.6% vs. 52.6%) of B-mode US for both observers compared with B-mode US alone (all, P=0.001).
    Conclusion: The VI of SMI for real-time breast US is highly reproducible and leads to improved diagnostic performance for differentiating between benign and malignant breast lesions in combination with B-mode US.

    더보기

    참고문헌 (Reference)

    1 Cohen J, "Weighted kappa: nominal scale agreement with provision for scaled disagreement or partial credit" 70 : 213-220, 1968

    2 Zhang XY, "Vascular index measured by smart 3-D superb microvascular imaging can help to differentiate malignant and benign breast lesion" 11 : 5481-5487, 2019

    3 박아영, "Up-to-date Doppler techniques for breast tumor vascularity: superb microvascular imaging and contrast-enhanced ultrasound" 대한초음파의학회 37 (37): 98-106, 2018

    4 Yadav L, "Tumour angiogenesis and angiogenic inhibitors: a review" 9 : XE01-XE05, 2015

    5 Park AY, "The utility of ultrasound superb microvascular imaging for evaluation of breast tumour vascularity: comparison with colour and power Doppler imaging regarding diagnostic performance" 73 : 304-311, 2018

    6 Ma Y, "The diagnostic value of superb microvascular imaging (SMI) in detecting blood flow signals of breast lesions: a preliminary study comparing SMI to color Doppler flow imaging" 94 : e1502-, 2015

    7 Zhan J, "Superb microvascular imaging: a new vascular detecting ultrasonographic technique for avascular breast masses: a preliminary study" 85 : 915-921, 2016

    8 Lee SW, "Role of color and power doppler imaging in differentiating between malignant and benign solid breast masses" 30 : 459-464, 2002

    9 Watanabe T, "Multicenter prospective study of color Doppler ultrasound for breast masses: utility of our color Doppler method" 45 : 1367-1379, 2019

    10 Hudson JM, "Inter- and intraoperator reliability and repeatability of shear wave elastography in the liver: a study in healthy volunteers" 39 : 950-955, 2013

    1 Cohen J, "Weighted kappa: nominal scale agreement with provision for scaled disagreement or partial credit" 70 : 213-220, 1968

    2 Zhang XY, "Vascular index measured by smart 3-D superb microvascular imaging can help to differentiate malignant and benign breast lesion" 11 : 5481-5487, 2019

    3 박아영, "Up-to-date Doppler techniques for breast tumor vascularity: superb microvascular imaging and contrast-enhanced ultrasound" 대한초음파의학회 37 (37): 98-106, 2018

    4 Yadav L, "Tumour angiogenesis and angiogenic inhibitors: a review" 9 : XE01-XE05, 2015

    5 Park AY, "The utility of ultrasound superb microvascular imaging for evaluation of breast tumour vascularity: comparison with colour and power Doppler imaging regarding diagnostic performance" 73 : 304-311, 2018

    6 Ma Y, "The diagnostic value of superb microvascular imaging (SMI) in detecting blood flow signals of breast lesions: a preliminary study comparing SMI to color Doppler flow imaging" 94 : e1502-, 2015

    7 Zhan J, "Superb microvascular imaging: a new vascular detecting ultrasonographic technique for avascular breast masses: a preliminary study" 85 : 915-921, 2016

    8 Lee SW, "Role of color and power doppler imaging in differentiating between malignant and benign solid breast masses" 30 : 459-464, 2002

    9 Watanabe T, "Multicenter prospective study of color Doppler ultrasound for breast masses: utility of our color Doppler method" 45 : 1367-1379, 2019

    10 Hudson JM, "Inter- and intraoperator reliability and repeatability of shear wave elastography in the liver: a study in healthy volunteers" 39 : 950-955, 2013

    11 Kook SH, "Evaluation of solid breast lesions with power Doppler sonography" 27 : 231-237, 1999

    12 Stanish WM, "Estimation of the intraclass correlation coefficient for the analysis of covariance model" 37 : 221-224, 1983

    13 Ma Y, "Correlation between superb microvascular imaging and pathological microvessel density in breast tumors" 1 : 30-, 2016

    14 Yang WT, "Correlation between color power Doppler sonographic measurement of breast tumor vasculature and immunohistochemical analysis of microvessel density for the quantitation of angiogenesis" 21 : 1227-1235, 2002

    15 Lee Eun Ji, "Combination of Quantitative Parameters of Shear Wave Elastography and Superb Microvascular Imaging to Evaluate Breast Masses" 대한영상의학회 21 (21): 1045-1054, 2020

    16 Giuseppetti GM, "Color Doppler sonography" 27 (27): S254-S258, 1998

    17 Lee WJ, "Breast cancer angiogenesis: a quantitative morphologic and Doppler imaging study" 2 : 246-251, 1995

    18 Yongfeng Z, "Application of a novel microvascular imaging technique in breast lesion evaluation" 42 : 2097-2105, 2016

    19 박아영, "An Innovative Ultrasound Technique for Evaluation of Tumor Vascularity in Breast Cancers: Superb Micro-Vascular Imaging" 한국유방암학회 19 (19): 210-213, 2016

    20 Chae EY, "Added value of the vascular index on superb microvascular imaging for the evaluation of breast masses: comparison with grayscale ultrasound" 40 : 715-723, 2021

    21 Mendelson EB, "ACR BI-RADS Atlas, Breast Imaging Reporting and Data System" American College of Radiology 1-173, 2013

    22 박아영, "A Prospective Study on the Value of Ultrasound Microflow Assessment to Distinguish Malignant from Benign Solid Breast Masses: Association between Ultrasound Parameters and Histologic Microvessel Densities" 대한영상의학회 20 (20): 759-772, 2019

    더보기

    분석정보

    View

    상세정보조회

    0

    Usage

    원문다운로드

    0

    대출신청

    0

    복사신청

    0

    EDDS신청

    0

    동일 주제 내 활용도 TOP

    더보기

    주제

    연도별 연구동향

    연도별 활용동향

    연관논문

    연구자 네트워크맵

    공동연구자 (7)

    유사연구자 (20) 활용도상위20명

    인용정보 인용지수 설명보기

    학술지 이력

    학술지 이력
    연월일 이력구분 이력상세 등재구분
    2023 평가 해외DB학술지평가 신청대상 (해외등재 학술지 평가)
    2020-01-01 등재 등재학술지 유지 (해외등재 학술지 평가) KCI등재
    2015-01-01 등재 등재학술지 유지 (등재유지) KCI등재
    2014-01-06 학술지명변경 한글명 : 대한초음파의학회지 -> ULTRASONOGRAPHY
    외국어명 : 미등록 -> ULTRASONOGRAPHY
    KCI등재
    2011-01-01 등재 등재 1차 FAIL (등재유지) KCI등재
    2009-01-01 등재 등재학술지 유지 (등재유지) KCI등재
    2006-04-10 학회명변경 영문명 : Korean Society Of Medical Ultrasound -> Korean Society of Ultrasound in Medicine KCI등재
    2006-01-01 등재 등재학술지 선정 (등재후보2차) KCI등재
    2005-01-01 등재 등재후보 1차 PASS (등재후보1차) KCI등재후보
    2003-01-01 등재 등재후보학술지 선정 (신규평가) KCI등재후보
    더보기

    학술지 인용정보

    학술지 인용정보
    기준연도 WOS-KCI 통합IF(2년) KCIF(2년) KCIF(3년)
    2016 0.33 0.33 0.23
    KCIF(4년) KCIF(5년) 중심성지수(3년) 즉시성지수
    0.17 0.13 0.599 0.18
    더보기

    이 자료와 함께 이용한 RISS 자료

    나만을 위한 추천자료

    해외이동버튼