<P><B>Abstract</B></P> <P>To avoid artifacts and interpretation mistakes, improvements in the uniformity of magnetic resonance imaging (MRI) is very important. The aim of this study is to develop a bias field-based unifo...
http://chineseinput.net/에서 pinyin(병음)방식으로 중국어를 변환할 수 있습니다.
변환된 중국어를 복사하여 사용하시면 됩니다.
https://www.riss.kr/link?id=A107446739
2019
-
SCI,SCIE,SCOPUS
학술저널
161-166(6쪽)
0
상세조회0
다운로드다국어 초록 (Multilingual Abstract)
<P><B>Abstract</B></P> <P>To avoid artifacts and interpretation mistakes, improvements in the uniformity of magnetic resonance imaging (MRI) is very important. The aim of this study is to develop a bias field-based unifo...
<P><B>Abstract</B></P> <P>To avoid artifacts and interpretation mistakes, improvements in the uniformity of magnetic resonance imaging (MRI) is very important. The aim of this study is to develop a bias field-based uniformity correction algorithm and to confirm the feasibility of its application with an integrated positron emission tomography (PET)/MRI system using standard MRI pulse sequences (T2-weighted, half-Fourier, single-shot turbo spin echo, T2-weighted turbo spin echo with fat suppression, and three-dimensional, volumetric, T1-weighted gradient recalled echo interpolated breath-hold examination using controlled aliasing in parallel imaging that results in increased acceleration). We used phantoms containing NaCl and NaCl + NiSO<SUB>4</SUB> solutions, and considered the percent image uniformity (PIU) as the evaluation parameter. The elicited PIU results using the correction algorithm were distinct for all sequences and phantoms. The results of this study suggested that the bias field-based algorithm allows uniformity improvements in MRI in terms of PIU.</P>