열형광선량계와 반도체검출기를 이용한 전신피부전자선조사의 선량평가
제영완,나경수,윤일규,박흥득,Je Young Wan,Na Keyung Su,Yoon IL Kyu,Park Heung Deuk 대한방사선치료학회 2005 대한방사선치료학회지 Vol.17 No.1
Purpose : To evaluate radiation dose and accuracy with TLD and diode detector when treat total skin with electron beam. Materials and Methods : Using Stanford Technique, we treated patient with Mycosis Fungoides. 6 MeV electron beam of LINAC was used and the SSD was 300 cm. Also, acrylic speller(0.8 cm) was used. The patient position was 6 types and the gantry angle was 64, 90 and $116^{\circ}$. The patient's skin dose and the output were detected 5 to 6 times with TLD and diode. Result : The deviations of dose detected with TLD from tumor dose were CA $+\;6\%$, thigh $+\;8\%$, umbilicus $+\;4\%$, calf $-\;8\%$, vertex $-\;74.4\%$, deep axillae $-\;10.2\%$, anus and testis $-\;87\%$, sole $-\;86\%$ and nails shielded with 4mm lead $+4\%$. The deviations of dose detected with diode were $-4.5\%{\sim}+5\%$ at the patient center and $-1.1\%{\sim}+1\%$ at the speller. Conclusion : The deviation of total skin dose was $+\;8\%{\sim}-\;8\%$ and that deviation was within the acceptable range(${\pm}\;10\%$). The boost dose was irradiated for the low dose areas(vertex, anus, sole). The electron beam output detected at the sootier was stable. It is thought that the deviation of dose at patient center detected with diode was induced by detection point and patient position.