김영범(Youngbum Kim),조한철(Hancheul Jo),김명관(Myungkwan Kim),김유현(Youhyun Kim) 고려대학교 보건과학연구소 2004 보건과학논집 Vol.30 No.1
Quality Assurance in the radiation therapeutic field is very important for the maintenance of the radiation therapy equipment as well as the safety of patient and therapist.<BR> Currently daily Quality Assurance study follows 26 daily Quality Assurance items according to recommendation in KFDA. But the daily Quality Assurance was hard to check various items by a restriction in time on a characteristic. In order to solve this problem, we prepared the new QA sheet for new QA equipment.<BR> We examined check time on 26 items by a conventional method over a week and measured check time which used new Quality Assurance equipment. When new equipment was used, check time was shortened, and we can include Output check measured in weekly Quality Assurance in new daily Quality Assurance sheet.<BR> If new daily Quality Assurance sheet be used, a check can do conventional daily QA including output check in 20 minutes.