윤석록,김명호,신동오,Yun, Suk-Rok,Kim, Myung-Ho,Shin, Dong-Oh 대한방사선치료학회 1987 大韓放射線治療技術學會誌 Vol.2 No.1
In the case of designing a high dose rate remote controlled afterloading treatment room with existing hospital facilities. We must construct the effective protective barriers so as to reduce the primary and scattered radiation up to the maximum permissible dose level. It is difficult to reinforce the barrier thickness of the shielding requirements because of the limited space and the problem of the existing building structure at the surrounding area. Therefore we can reduce the intensity of primary radiation to the required degree at the location of interest with installing the appropriate I shaped Pb barriers between the radiation source and the shielding wall of the concrete. As a result, it was possible to reduce the intensity of the primary radiation below the M.P.D level by using additional Pb barriers instead of increasing thickness of concrete wall.
코발트60 원격치료기에 대한 Semiuniversal Wedge Filter의 제작과 임상응용
김명호,윤석록,신동오,Kim, Myung-Ho,Yun, Suk-Rok,Shin, Dong-Oh 대한방사선치료학회 1989 大韓放射線治療技術學會誌 Vol.3 No.1
고 에너지 방사선으로 용적이 적은 표재성 종양 및 부정형의 체표면과 경사진 조사야에 대한 방사선 치료시 종양 용적내에 균등한 선량분포를 얻기 위해서 Wedge filter 기법이 많이 사용되고 있다. 본 연구의 목적은 코발트 $60{\Gamma}$선 치료를 위한 Semi-Universal Wedge filter의 제작과 임의의 조사야에 대한 최적한 출력 및 정확한 조작방법을 기술하고 방사선 치료계획의 최적화를 기하기 위해서 Wedge filter의 임상응용 방법을 제시하고자 한다. Wedge filter are used in high energy radiation therapy to obtain uniform isodose lines at small volume tumor and irregular body surface. The aim of this study is not only to describe the method of construction and setting of semi-universal wedge filter for $^{60}Co$ radiation therapy but also to get the optimum treatment planning by clinical application of wedge filter.